Selection
Guide: KT19-Series II- Infrared Radiation Pyrometer |
|
Click
here to download complete
KT19-Series II technical specifications, including
temperature ranges and resolution tables. |
|
Twenty spectral models are
available with the KT19-Series II. A wide
selection of lenses and detector sizes makes the
KT19-Series II the most versatile compact
instrument on the market. The following table
lists the temperature ranges and basic
applications associated with each model.
Technical
specifications for the KT19-Series II
follows below. |
Selection
Guide: Infrared Radiation Pyrometer KT19-Series
II
| Model |
Spectral Response/um |
tmin..tmax/°C |
Typical Application |
| KT19.01 |
2.00...2.70 |
250...2,000 |
metals, metal oxides, ceramics and glass
volume |
| KT19.02 |
2.00...4.50 |
100...1,200 |
metals, metal oxides, glass volume (lower
temperatures) |
| KT19.21 |
3.43±0.15 |
80...350 |
plasic film with CH-band, coating materials
(oil, pints) |
| KT19.23 |
6.80±0.15 |
40...400 |
thin film plastics, e.g. PE, PP, PVC |
| KT19.24 |
7.93±0.15 |
0...400 |
thin film plastics, e.g. PET, PA,
fluorcarbon |
| KT19.25 |
8.05±0.15 |
0...400 |
thin film plastics, e.g. PTFE, PET, PVC |
| KT19.41 |
3.90±0.10 |
200...2,500 |
glass volume, measurements through hot
gases and flames |
| KT19.42 |
4.90...5.50 |
100...2,500 |
glass (processing), quartz |
| KT19.43 |
7.50...8.20 |
0...2,500 |
glass (thin plates), quartz, ceramics |
| KT19.61 |
4.26±0.13 |
300...2,500 |
hot gases and flames (CO2-band) |
| KT19.62 |
4.50±0.10 |
300...2,500 |
hot gases and flames (CO2-and CO-band) |
| KT19.63 |
4.66±0.10 |
300...2,500 |
hot gases and flames (CO-band) |
| KT19.69 |
x...y |
400...2,500 |
hot gases in incinerators and fossil fuel
fired utility boilers, rotary kiln |
KT19.81
KT19.82
KT19.83 |
8...10
8...14
8...20 |
0...1,000
-50...1,000
-50...1,000 |
paper, textiles, rubber, wood, ceramics,
thicker plastics (>1mm),painted or coated
surfaces, asphalt,building materials, electronic
components, food, liquids |
| KT19.85 |
9.6-11.5 |
-50
200 |
Meteorology, Long distance measurement
water , Ice & cloud temperature |
| KT19.xx |
|
|
several spectral response bands possible,
temperature range depending on application |
|
KT19-II-Series
General Specifications
| Temperature range |
Depends on model, minimum and maximum
measuring temperature; see table above |
| Temperature resolution (NETD)* |
Depends on model, detector size, measuring
temperature and response time; typical value ±0.1°C |
| Accuracy |
±0.5°C plus 0.7% of the
difference between target temperature and
housing temperature |
| Long-term stability |
Better than 0.01% of the absolute measured
temperature in Kelvin/month |
| Lenses* |
Several far-focus and close-focus lenses
with different optical characteristics are
available, including vacuum tight lens barrels |
| Field of view* |
Depends on model, detector and lens type.
For close-focus lenses: from 0.8 mm diameter at
>20 mm distance; for far-focus lenses:
typically 20 to 30 mm diameter at distances >1,000
mm. |
| Aiming options |
Optional thru-the-lens sighting with
integrated temperature indicator; alternatively
built-in laser pointer; or focus laser. Other
optional aiming aids (optical or mechanical) |
| LCD display on rear panel |
Indicating measured temperature,
emissivity, high/low alarm, programming menu |
| Keyboard on rear panelfor programming of
|
Emissivity, ambient temperature
compensation, analog output, response time,
temperature unit, serial interface, read-out of
miximum/minimum value, calibration procedure,
on-/off-switch for laser pointer |
| Emissivity setting |
Adjustable from 0.1 to 1.0 in 0.001
increments.Reflectivity and Transmissivity
compensation |
| Response time |
Selectable: 5mS, 10mS, 30mS, 100mS, 300mS,
1S, 3S, 10S, 30S, 60S, 120S, 240S, 360S, 480S,
600S for +90 |
| Analog output |
4 scalable output signals: linear voltage
or current, 0-1(10)V or 0(4)-20 mA. Integrated
peak/valley picking w/ programmable decay.
Output can be temperature or relative radiance |
| Serial interface |
RS-232 interface, bi-directional, baud rate
9600 to 115,000 bps, for free programming and
data transfer |
| Alarm relays |
Two set-points indicated on LCD display |
| Thermal switch |
Monitoring housing temperatue |
| Power requirements |
22...30 VDC or 24 VAC ±10%, 48...400
Hz; <150 mA with 24 VDC |
| Permissible ambient temperature |
-20°C...60°C, w/ WK300 cooling
jacket: 300°C |
| Storage temperature |
-20°C...70°C |
| Housing protection/Weight |
IP 65 (DIN 4005) NEMA 4 equivalent/approx.
1.5 kg |
|
| Wintronics
Inc., exclusive US agent for Heitronics Infrared
Radiation Pyrometers |
|
|
|
|
|