Temperature range |
0°C ... 900°C |
Temperature resolution (NETD) |
Depends on measuring temperature and response time, typical value 0.2°C (at 300 ms, 100 °C, e = 1) |
Accuracy (uncertainty) |
± 0.8°C plus 0.8% of the difference between target and sensor head temperature |
Long term stability |
Better than 0.01% of the absolute measured temperature per month |
Field of view diameter |
From Ø 1 mm, depends on lens |
Field of view marking |
Laser pointer as accessory for non transparent lenses
Pilot laser, built-in: aims the center of the field of view in any distance
(More mechanical pointers are available) |
Laser function |
Time out or permanent operation, while flashing or continuous marking |
Spectral ranges |
2 - 2.7 µm, 3.9 µm, 4.9 - 5.6 µm, 7.93 µm, 7.5 - 8.2 µm, 8 - 14 µm |
Programmable functions via serial interface |
Emissivity, Environmental temperature, Analog output, Function of analog output, Response time, Temperature unit, Valley/ Peak-picker with decay function, Laser operation |
Emissivity |
0.100 to 1,000 in 0,001-steps |
Response time |
from 30 ms to 10 s (0,03, 0,1, 0,3, 1, 3, 10 s); Option: 5 ms available |
Temperature unit |
°C, °F or K |
Analog output (Hardware)
Analog output (Functions)
|
linear 0-20mA or 4-20mA, Zoom function for temperature span > 50 °C.
Actual value, Maximum value or Minimum value
|
Serial interface |
RS232-Interface, bi directional, 9,600 to 57,600 bps, for programming and data transfer |
Thermal switch |
Monitors the instrument temperature |
Power requirements |
10.5-32 VDC or 10.8 VAC to 26.4 VAC, 48-400 Hz
Approx. 2.5 W |
Permissible operating temperature |
-25°C ... 60°C
With protective and cooling housing WK15 up to 300 °C
|
Storage temperature |
-40°C ... 85°C |
Protective class, Weight |
IP68 (IEC), NEMA 6; 1.3 kg |
Housing |
Stainless steel |
Wintronics Inc., exclusive US agent for HEITRONICS Infrared Radiation Pyrometers